Hach Master Catalog 2025
Ammonia Monochloramine • Process Instruments 5500sc Ammonia Monochloramine Analyzer Chloramination made easy.
81
The Hach 5500sc Ammonia Monochloramine Analyzer provides all the information you need to eliminate nitri fi cation events and taste and odor issues, giving you total con fi dence in your process. Easy operation The analyzer o ff ers an easy-to-operate, low-maintenance solution to help you maintain the proper chlorine to ammonia ratio and assure there is no free ammonia in your system that could lead to nitri fi cation. We removed the traditional pumps associated with reagent delivery and replaced them with a pressurized system to eliminate the hassles and maintenance that pumps can cause. Reliability without all the work With Prognosys, the analyzer’s onboard predictive diagnostic software, you will have early insight into the measurement reliability and service requirements of your instruments. A user-friendly interface, color coded reagent bottles and at-a-glance status lights will o ff er peace of mind the instrument is up and running. Improved control of your chloramination process
Technical Data* Range
Order Information 5500.AMC.1.KTO 5500sc Ammonia Monochloramine Analyzer, 1 Channel 5500.AMC.2.KTO 5500sc Ammonia Monochloramine Analyzer, 2 Channel 5500.AMC.3.KTO 5500sc Ammonia Monochloramine Analyzer, 1 Channel with External Filter 5500.AMC.4.KTO 5500sc Ammonia Monochloramine Analyzer, 2 Channel with External Filter The 5500sc Ammonia Monochloramine Analyzer provides continuous monitoring, so you can trust that your instrument is giving you the information you need to control your chloramination process.
0.01 - 2 ppm (as N), 0.05 - 10.0 mg/L (as Cl 2 )
Total Ammonia, Monochloramine, and Free Ammonia 1 or 2 in programmable sequence
Parameter
Number of sample streams
± 5% or 0.01 ppm (as N) for 5 - 40 °C; ±10% or 0.02 ppm for 40 - 50 °C, whichever is greater 3% or 0.01 ppm (as N), whichever is greater
Accuracy
Repeatability
Lower Limit of Detection (LOD)
0.01 ppm (as N) Within 5 minutes
Response Time
1L of reagents every month with 5 minute cycle time
Reagent Consumption
Operating Temperature Range 5 - 45 °C (41 - 113 °F) Operating Humidity 5 - 95% non-condensing Sample Pressure 0.17 - 6.8 bar (2.5 - 87 psi) Sample Temperature 5 - 50 °C (41 - 122 °F) Sample Flow Rate 100 - 1000 mL/minute Grab Sample
Reagents 25233000 25234000 25235000 25236000 25237000 25238000 25239000 9560501 9560502 25224000 6792500 25277000 Accessories 9179700
5500sc Ammonia Monochloramine Analyzer Reagent Set 5500sc Ammonia Monochloramine Reagent 1, 1L 5500sc Ammonia Monochloramine Reagent 2, 1L 5500sc Ammonia Monochloramine Reagent 3, 1L 5500sc Ammonia Monochloramine Standard 1 (0 ppm NH 3 ), 2L 5500sc Ammonia Monochloramine Standard 2 (2 ppm NH 3 ), 2L 5500sc Ammonia Monochloramine Acidic Surfactant Wash, 2 L
Grab Sample In and Grab Sample Out capability
Mounting
Wall, panel, or table
6mm OD push-to-connect fi tting for plastic tubing 11mm (7/16 in.) ID slip-on fi tting Four 0/4-20 mA Outputs, load impedance 600 ohm maximum 4-20mA Output, Optional: Modbus RS485, Pro fi bus DP with external controller 804 x 452 x 360 mm (31.65 x 17.8 x 14.17 in.) CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 cETLus (UL 61010-1:2012; NEMA 250:2003; CSA C22.2 No 61010-1:2012) Australian RCM Marking 20.5 kg (45.5 lb) without reagents and standards, 30 kg (66.2 lb) with reagents, standards and cleaning solution IP56 / NEMA 4X
Inlet
Outlet
Outputs
Power cord, North American
5500sc Ammonia Monochloramine Maintenance Kit, 1 Channel 5500sc Ammonia Monochloramine Maintenance Kit, 2 Channel Colorimeter Cleaning Kit for 5500sc Ammonia Monochloramine Analyzer
Communication
Dimensions (H x W x D)
Enclosure Rating
5500sc Modbus Module Kit
Prognosys License for the 5500sc Ammonia Monochloramine Analyzer
Certi fi cations
LearnMore
Weight
*Subject to change without notice.
Made with FlippingBook - Online catalogs